Prodigy Advance - GE Healthcare

Prodigy Advance - GE Healthcare

Prodigy Advance Tailored for advanced skeletal and metabolic health assessment Optimal health depends on accurate diagnosis and preventive treatment. ...

460KB Sizes 3 Downloads 47 Views

Recommend Documents

prodigy - GE Healthcare
For Today – and for the Future. Prodigy Pro is for customers with a vision of the future – a fan-beam platform that

Prodigy Pro Spec Sheet - GE Healthcare
The Prodigy® Pro is designed to look beyond bone mineral density. Utilizing a narrow-angle fan beam coupled with a dire

Lunar Prodigy Primo - GE Healthcare India
Indications for use: The Prodigy series bone densi- tometer provides an estimate of bone mineral density and fat and lea

Cross-Calibration of GE Healthcare Lunar Prodigy and iDXA Dual
Mar 27, 2016 - measurements with the Prodigy and the iDXA scanners at the same visit. The bone mineral density (BMD) mea

MRC Prodigy Advance User's Manual
The Prodigy Advance does not support NON-decoder equipped or analog locos. MRC's earlier DCC systems do support running

Advance Directive - Covenant HealthCare
Courtesy of Covenant HealthCare, Saginaw, MI • DPOA-HC .... If you do not personally know the person signing this docu

“PRODIGY ADVANCE, ADVANCE SQUARED, AND - Horizon Hobby
When you get your new Prodigy Advance or Express DCC system up and running for the first time you should clear you handh

Review of MRC Prodigy Advance DCC System
MRC has released a new DCC control system named the Prodigy Advance. ... of three units: an AC wall transformer, the DCC

mrc prodigy advance - MTH Electric Trains
SHORT ADDRESS CHANGES. To change an M.T.H. HO locomotive's address on the Prodigy Advance system you'll need to use Prog

imagination at work - GE Healthcare
GE Healthcare provides transformational medical technologies and services ... *GE, the GE Monogram, Centricity and imagi

Prodigy Advance Tailored for advanced skeletal and metabolic health assessment Optimal health depends on accurate diagnosis and preventive treatment. That’s why so many ™ professionals around the globe rely on Prodigy for comprehensive body composition analysis, including bone mineral density (BMD) and lean and fat tissue mass.

Prodigy delivers reliable dual-energy X-ray absorptiometry (DXA) with excellent precision and extremely low radiation dose. Prodigy efficiently streamlines patient care and practice workflow. You can trust Prodigy to help ensure the vitality of your patients and your practice.

Scanner dimensions Full-size bed:

Compact bed: 67cm (26”) 128cm (51”) 25cm (10”) 63cm (25”)

262cm (103”)

201cm (79”)

109cm (43”)

US/CALA version - enCORE v16 - Some features may not be available in all markets.

Prodigy Advance specifications (nominal)

Scanner table specifications: Scanner size (full-size bed) . . . . . . . . . . . . . . 262(W) x 109(D) x 128(H)cm (103” x 43” x 51”) Scanner size (compact bed) . . . . . . . . . . . . . 201(W) x 109(D) x 128(H)cm (79” x 43” x 51”) Scanner weight (full-size bed) . . . . . . . . . . . . . . . . . . . . . . . . 272kg (599lbs) Scanner weight (compact bed) . . . . . . . . . . . . . . . . . . . . . . . 254kg (559lbs) Patient table top height (adjustable) . . . . . . . . . . . . . . . . . . . . >63cm (25”) Maximum patient weight supported . . . . . . . . . . . . . . . . 159kg (350 Ibs) Drive system . . . . . . . . . . . . . . stepper motor with reinforced drive belts Active scan area (full-size bed) . . . . . . . . . . . . . . . . . . . 197.5cm x 60.0cm Active scan area (compact bed) . . . . . . . . . . . . . . . . . . 135.2cm x 59.5cm Start position indicator . . . . . . . . cross laser light (class II, <1mW power) Pad . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . washable patient mat Attenuation of patient support table . . . . . . . . . . . . . . . . . . . . <0.7mm AL Communication cable . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7.62m (25ft) serial Scanner leakage current . . . . . . . meets IEC 60601-1 safety standard Detector specifications: Detector . . . . . . . . . . . . . . . . . . . . . . . . . . . . . LYSO x-ray counting detector Computer specifications: Non-US customers will need to verify that the computer is certified to local requirements. The computer must meet the minimum requirements that follow: • 1.2 GHz Intel™ Celeron™ • External hard drive or 2.69 GHz AMD Athlon™ (data archive location) processor • Internet Explorer® version 8.0 • Windows® 7 Professional • Windows-compatible printer (32-bit, 64-bit) • 1 Serial port required • 2 GB RAM • Adobe™ Acrobat™ reader • 160 GB hard disk • Serial Port: Onboard RS-232 • DVD-R drive 115K baud DB-9 or StarTech • 17” SVGA monitor with at PCI serial adapter PCI1S950DV least 1024 x 768 32-bit color (GE Healthcare P/N LU44192)

Environmental specifications

Minimum room dimensions5 Full-size bed:

Compact bed:

2.4m (8')

Power . . . . . . . . . . . . . . . . . 100-120 VAC 50/60Hz 20A dedicated circuit 220-240 VAC 50/60Hz 10A dedicated circuit Consumption . . . . . . . . . . . . . . . . . . . . . . . . . Idling 40VA, Scanning 450VA Distortion . . . . . . . . . . . . . . . . . . . sinusodal waveform, less than 5% THD Humidity . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20%-80% non-condensing Room temperature . . . . . . . . . . . . . . . . . . . . . . . . . . . 18°C-27°C (65°F-81°F) Scanner heat output . . . . . . . idling 150 BTU/hr, scanning 1500BTU/hr Console heat output . . . . . . . . . . . . approx. 400BTU/hr with 17” monitor Ventilation . . . . . . . . . . . . . . . . all cooling vents must remain unblocked Dust, fumes, debris . . . . . . . . . . . install system in clean, ventilated area

2.4m (8')

Software applications and features: OPTIONAL Clinical applications: Workflow: • AP spine • Previous scan image • Femur comparison • DualFemur • OneVision • Forearm/supine forearm • Automatic metal detection • Total body BMD • Image preview • Dual-energy Vertebral • OneScan measurement Assessment (DVA) • SmartScan (lateral and AP) • QuickView measurement • Fracture risk assessment tool: (10sec) FRAX®2 Analysis & reporting: • Total and regional body • Custom region of interest composition1 analysis • Advanced body composition • Composer reporting tools (data visualization, trending • Custom reference creation & reporting tools)1 • ScanCheck • Advanced Hip Assessment • Practice management tools (AHA) with hip strength Connectivity: analysis • HIPAA secure view • Pediatric spine/femur/ • SQL server 1 Complete Pediatric • DICOM interface Hand/supine hand • HL7 interface3 2 Small animal total body TeleDensitometry (e-mail, fax)4 • CoreScan™ (visceral fat Multi-User DataBase access quantification)1 (MUDB) (1-3 users) • Orthopedic hip Multi-User DataBase access • Orthopedic knee (MUDB) (1-10 users)

3.0m (10')

2.3m (7.5')

About GE Healthcare GE Healthcare provides transformational medical technologies and services that are shaping a new age of patient care. Our broad expertise in medical imaging and information technologies, medical diagnostics, patient monitoring systems, drug discovery, biopharmaceutical manufacturing technologies, performance improvement and performance solutions services help our customers to deliver better care to more people around the world at a lower cost. In addition, we partner with healthcare leaders, striving to leverage the global policy change necessary to implement a successful shift to sustainable healthcare systems. Our “healthymagination” vision for the future invites the world to join us on our journey as we continuously develop innovations focused on reducing costs, increasing access and improving quality around the world. Headquartered in the United Kingdom, GE Healthcare is a unit of General Electric Company (NYSE: GE). Worldwide, GE Healthcare employees are committed to serving healthcare professionals and their patients in more than 100 countries. For more information about GE Healthcare, visit our website at www.gehealthcare.com. 1. Full-size table only. 2. Laboratory animals only 3. Standard in US, may be optional in other regions. 4. Additional hardware may be required for fax capabilities. 5. A small room kit with isolation transformer may be required. Please refer to local regulations. © 2015 General Electric Company – All rights reserved. GE, GE monogram, Prodigy and CoreScan are trademarks of General Electric Company or one of its subsidiaries. FRAX is a registered trademark of the World Health Organization Collaborating Centre for Metabolic Bone Diseases, University of Sheffield, UK. DICOM is a trademark of National Electrical Manufacturers Association. Windows and Internet Explorer are registered trademarks of Microsoft Corporation. Intel and Celeron are trademarks of Intel Corporation. Athlon is a trademark of Advanced Micro Devices, Inc. Adobe and Acrobat are trademarks of Adobe Systems Incorporated. All other trademarks, service marks, company names and product names are the property of their respective owners. General Electric Company reserves the right to make changes in specifications and features shown herein, or discontinue the product described at any time without notice or obligation. Contact your GE representative for the most current information. GE Healthcare, a division of General Electric Company Indications for use: The Prodigy series bone densitometer provides an estimate of bone mineral density and fat and lean tissue mass. The values can then be compared to a reference population at the sole discretion of the physician. CAUTION: Federal Law restricts this device to sale by or on the order of a physician.

Lunar Product Division Americas GE Healthcare Lunar Global Headquarters P.O. Box 7550 Madison, WI 53707-7550 T: +1-800-535-7339 F: +1-608-223-2482 Lunar Product Division Europe GE Healthcare Europe Buc Headquarters 283 Rue de la Minière 78533 BUC Cèdex, France T: +33-800-90-87-19 Asia & Pacific GE Healthcare Lunar 3/F GE China Technology Park No. 1 Hua Tuo Road Shanghai 201203, China T: +86-21-38777888 (Ext. 60128 or 60480) F: +86-21-38777451 Lunar Product Division EAGM (ME, Africa, Russia CIS, Turkey CA) GE Healthcare Building #18 Dubai Internet City P.O. Box 74594 Dubai, United Arab Emirates T: +971-4-429-6218 F: +971-4-429-6218

US/CALA version - enCORE v16 - Some features may not be available in all markets. JB20792XXi